Gas Mixing System GMS 150 can produce precise mixtures of up to 4 different gases. The flows of the individual input gases are measured by thermal mass flow meters and adjusted by integrated mass flow controllers.
GMS 150 can also be coupled with the PSI Photobioreactors, Multi-Cultivators, or FytoScopes. Standard version of the device designed for laboratory scale systems (PBR FMT 150 and MC-1000-OD) enables the maximum total flow of 2 l/min and maximum CO2 flow rate 40 ml/min. The typical GMS version designed for large-scale facilities (e.g. PSI large-scale Photobioreactors and FytoScopes) allows the maximum total flow 10 or 20 l/min for channel 1 and 400 or 800 ml/min for channel 2.
Newly, PSI introduces Gas Mixing System GMS 150-MICRO - friendly-priced basic version of the GMS 150. It retains all features of the GMS 150 and the major difference is the limited choice of gas flow meters restricted to maximum total flows ranging from 10 ml/min to 2.5 l/min and lower accuracy of mixing. Standard version of the GMS 150-MICRO is equipped with two channels (channel 1 for Air-N2; channel 2 for CO2); the maximum total flow is 2 l/min (channel 1), the maximum flow rate of channel 2 is 40 ml/min. Optionally the system can be upgraded to four channels version according to customer requirements. Preferably, the GMS 150-MICRO is intended for use with Multi-Cultivators and standard laboratory versions of Photobioreactors (FMT 150/400 and FMT 150/1000).
Accuracy (Incl. Linearity):
±0.5% Rd plus ±0.1% FS (±1% FS for ranges 3 – 5 ml/min; ±2% FS for ranges < 3 ml/min)
Minimum Possible Flow for Each Channel:
2% of total flow
Flow Ranges:
Max. total flow rate: 2 or 10 or 20 l/min typically (customized max. flow rate between 1 ml/min – 1,000 l/min)
Min. flow rate: 2% of the max. flow rate
Media:
Dry, clean, non-explosive and non-corrosive gasses
Control Stability:
< ±0.1% FS (typical for 1 l/min N2)
Setting Time:
1 – 2 seconds
Warm-Up Time:
30 min for optimum accuracy, 2 min for accuracy ± 2% FS
Temperature Sensitivity:
zero: < 0.05% FS/°C; span: < 0.05% FS/°C
Pressure Sensitivity:
0.1%/bar typical N2
Attitude Sensitivity:
max. error at 90° off horizontal 0.2 % at 1 bar, typical N2
Input Pressure:
3 bar – 5 bar
Operating Temperature:
15 – 50 °C
Input / Output Connectors:
Parker Prestolok (6 mm)
Seals:
Viton
Display:
8 × 21 characters LC display
Dimension:
37 × 28 × 15 cm
Weight:
7 kg
Power Supply:
115 – 230 VAC
Accuracy (Incl. Linearity):
±1.5 % Rd plus ±0.5 % FS
Minimum Possible Flow for Each Channel:
2 % of total flow
Flow Ranges:
Max. total flow rate: 2 l/min typically (customized max. flow rate between 10 ml/min – 2,5 l/min)
Min. flow rate: 2% of the max. flow rate
Media:
Dry, clean, non-explosive and non-corrosive gasses
Repeatability:
For flows <20 ml/min: ±0.5 % FS; for flows >20 ml/min: ±0.5 % RD
Setting Time:
1 second
Temperature Sensitivity:
zero: <0.01 % FS/°C; span: < 0.02 % FS/°C
Attitude Sensitivity:
max. error at 90° off horizontal 0.5 % at 1 bar, typical N2
Input Pressure:
3…5 bar (maximum operation pressure 8 bar)
Operating Temperature:
15 - 30 °C Input / Output Connectors: Parker Prestolok (6 mm)
Seals:
Viton
Display:
8 × 21 characters LC display • Dimension: 37 cm× 28 cm× 5 cm
Weight:
5 kg
Power Supply:
115-230 VAC
Version | Characteristics | Price |
---|---|---|
Gas Mixing System GMS 150 |
High-precision gas mixing system includes two channels: channel 1 for air-N2: 40 - 2,000 ml/min channel 2 for CO2: 0.8 - 40 ml/min Maximum total flow is 2 L/min. |
9,390 € |
Gas Mixing System GMS 150-MICRO |
GMS 150-MICRO serves for precise gas mixing. It includes two channels: channel 1 for air-N2: 40 - 2,000 ml/min channel 2 for CO2: 0.8 - 40 ml/min Maximum total flow is 2L/min. |
6,490 € |
Gas Mixing System GMS 150/TORA-10L |
High-precision gas mixing system includes two channels:
channel 1 for air-N2: 200 - 10,000 ml/min channel 2 for CO2: 8 - 400 ml/min Maximum total flow is 10 L/min (for 25L PBR). |
9,390 € |
Gas Mixing System GMS 150/TORA-20L |
High-precision gas mixing system includes two channels: channel 1 for air-N2: 400 - 20,000 ml/min channel 2 for CO2: 16 - 800 ml/min Maximum total flow is 20 L/min (for 100L PBR). |
9,390 € |
Gas Mixing System GMS 150/CUST |
High-precision gas mixing system includes two channels: channel 1 for air-N2; channel 2 for CO2. Maximum total flow is variable according customer needs. |
Enquire for price |
Optional Accessories
Additional Channel to Gas Mixing System GMS 150 * | One additional channel. | 3,780 € |
Additional Channel to Gas Mixing System GMS 150-MICRO * | One additional channel. | 2,590 € |
GMS 150 Valve Recalibration * | Recalibration of one valve to different flow range or different gas. The possibility of such recalibration must be consulted with PSI prior to the order. | Enquire for price |
Regular Preventive Maintenance of the GMS 150 or GMS 150-MICRO (2-Channel System) * | Regular general maintenance is recommended every two years. It includes: inspection, cleaning, and recalibration of the whole system (additional surcharge may apply if replacement of faulty parts is needed). | 1,059 € |