Spectrometer SM 9000 is used for high-resolution and high-sensitivity measurements in fibre optic spectroscopy applications. It covers UV/VIS or NIR spectral range. SM 9000 can be used as a stand-alone instrument or as a optional enhancement of other PSI instruments for precise analysis of the spectral curves (in irradiance, absorption, emission mode).
Entrance Slit:
70 ×1,400 µm (optical entrance)
Grating:
Flat-field correction
Spectral Range:
200 – 980 nm
Wavelength Accuracy Absolute:
< 0.5 nm
Reproducibility:
< 0.1 nm
Temperature – Induced Drift:
< 0.01 nm/K
Spectral Distance of Pixel:
Δλ pixel ≈ 0.8 nm
FWHM (Full Width Half Maximum):
Δλ < 3 – 4 nm (UV-NIR Version)
Straylight:
0.1 % measured at 340 nm with deuterium lamp (transmission of NaNO2 solution, 50 g/l, 1 cm)
CCD Array:
Thermoelectrically cooled Hamamatsu S 7031
Number of Pixels:
1,044 × 64
Dimensions of Pixels:
24 × 24 mm2
System Data:
16 Bit A/D conversion
Noise:
2 – 4 count standard deviation
Version | Characteristics | Price |
---|---|---|
Spectrometer SM 9000 | Allows applications in the UV/VIS to NIR range. Includes spectra-processing firmware. | 16,189 € |